Parameters:
Applications
It is suitable for the magnetic measurement of rare earth permanent magnets (Nd-Fe-B, Sm-Co alloys) and general permanent magnets (ferrite, Al-Ni-Co alloys). The magnetic parameters, such as Br, Hcb, Hcj, (BH) m, Hk, Hk/Hcj, can be directly obtained by measuring the demagnetization curve and hysteresis loop from room temperature to 200 °C.
Features
l The magnetic field is measured by H-coil or Hall effect
l The measurement process is quasi-static, which avoids eddy current field, eliminates the phase angle between magnetic field and magnetic polarization intensity, and makes the sample magnetized uniformly
l The electromagnet has advanced magnetic circuit design and high excitation field
l The measuring system can work continuously
l The high temperature measurement has fast heating rate and temperature control accuracy of ±0.3 °C
Parameters
l Temperature range: room temperature ~ 200 °C
l The maximum magnetic field: 2.6 T
l Accuracy of magnetic flowmeter: 0.2%
l Accuracy of gaussmeter: 0.2%